Description
CO2 Incubator ICO240
Temperature
Working-temperature range | 5 °C above ambient temperature up to 50 °C Standard sterilization programme: 60 minutes at 180°C (without removing the sensors) |
---|---|
Setting temperature range | 18 to 50 °C |
Setting accuracy temperature | 0.1 °C |
Temperature | 2 Pt100 sensors Class A in 4-wire-circuit, mutually monitoring and taking over the performance at the same temperature value |
Humidity
Setting range active humidity control | 40 to 97 % rh and rh-Off |
---|---|
Setting accuracy humidity | 0.5 % rh |
Humidity control (standard) | Humidity limitation thanks to a Peltier element; when water dish is full and inserted, the Peltier element limits the value of relative humidity in the interior to 93 % rh /- 2.5 % |
Addtional humidity control (standard) | Capacitive humidity sensor for measuring and displaying the relative humidity (regular Option K6 – free of charge until 31st Dec 2017) |
Control of standard components
CO2 control | Digital electronic CO2 control with dual beam NDIR system, with auto-diagnostic system and acoustic fault indication, barometric pressure compensation |
---|---|
Adjustment range CO2 | 0 to 20 % CO2 |
Setting accuracy CO2 | 0,1% |
Variation in time CO2 | /- 0.2 % CO2 |
Adjustment range O2 | 1 to 20 % O2 |
Setting accuracy O2 | 0.1 % O2 |
Control technology
ControlCOCKPIT | TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays. |
---|---|
Language setting | German, English, Spanish, French, Polish, Czech, Hungarian |
Function SetpointWAIT | the process time does not start until the set temperature is reached |
adjustable parameters | temperature (Celsius or Fahrenheit), fan speed, CO2, programme time, time zones, summertime/wintertime |
Communication
Interface | Ethernet LAN, USB |
---|---|
Documentation | programme stored in case of power failure |
Programming | AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port |
Safety
Autodiagnostic system | integral fault diagnostics for temperature, CO2 and humidity limit control |
---|---|
Alarm | visual and acoustic |
Heating concept
6 sides | large-area multi-function heating system on four sides with additional door and back heating to avoid condensation |
---|
Standard equipment
Internals | 2 perforated stainless steel shelf/shelves |
---|---|
Internals | 1 stainless steel water dish |
Door | fully insulated stainless steel door with 2-point locking (compression door lock) |
Door | inner glass door with opening (8 mm Ø) to take gas sample |
Stainless steel interior
Interior | material 1.4301 (ASTM 304), corrosion resistant |
---|---|
Volume | 241 l |
Dimensions W x H x D in mm | w(A) x h(B) x d(C): 600 x 810 x 500 mm |
Max. number of internals | 12 |
Max. loading of chamber | 140 kg |
Max. loading per internal | 15 kg |
Textured stainless steel casing
Dimensions | w(D) x h(E) x d(F): 759 x 1176 x 691 mm |
---|---|
Housing | rear zinc-plated steel |
Electrical data
Voltage Electrical load | 230 V, 50/60 Hz approx. 2000 W |
---|---|
Voltage Electrical load | 115 V, 50/60 Hz approx. 2000 W |
Ambient conditions
Set Up | The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm. |
---|---|
Ambient temperature | 10 °C to 35 °C |
Humidity rh | max. 70 %, non-condensing |
Altitude of installation | max. 2,000 m above sea level |
Overvoltage category | II |
Pollution degree | 2 |
Packing/shipping data
Transport information | The appliances must be transported upright |
---|---|
Customs tariff number | 8419 8998 |
Country of origin | Federal Republic of Germany |
WEEE-Reg.-No. | DE 66812464 |
Dimensions approx incl. carton | B x H x T: 840 x 1360 x 900 mm |
Net weight | approx. 110 kg |
Gross weight carton | approx. 145 kg |