Peltier-Cooled Incubator IPP750ecoPlus
Temperature
Working temperature range | from 0 (at least 20 below ambient temperature) to +70°C |
---|---|
Setting accuracy temperature | 0.1°C |
Temperature | 2 Pt100 sensors Class A in 4-wire-circuit, mutually monitoring and taking over the performance at the same temperature value |
Control technology
ControlCOCKPIT | TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-color displays. |
---|---|
Language setting | German, English, Spanish, French, Polish, Czech, Hungarian |
Timer | Digital backwards counter with target time setting, adjustable from 1 minute to 99 days |
Function HeatBALANCE | adapting the distribution of the heating performance of the upper and lower heating circuit from -50 % to 50 % |
Function SetpointWAIT | the process time does not start until the set temperature is reached |
Calibration | three freely selectable temperature values |
adjustable parameters | temperature (Celsius or Fahrenheit), program time, time zones, summertime/wintertime |
Ventilation
Convection | forced ventilation by Peltier fan |
---|
Communication
Documentation | program stored in case of power failure |
---|---|
Programming | AtmoCONTROL software on a USB stick for programming, managing and transferring programs via Ethernet interface or USB port |
Safety
Temperature control | over- and under temperature monitor TWW, protection class 3.3 or adjustable temperature limiter TWB, protection class 2, selectable on display |
---|---|
AutoSAFETY | additionally integrated over- and under temperature protection “ASF”, automatically following the setpoint value at a preset tolerance range, alarm in case of over- or under temperature, heating function is switched off in case of overtemperature, cooling function in case of under temperature |
Autodiagnostic system | for fault analysis |
Alarm | visual and acoustic |
Heating concept
Peltier | Heating and cooling performance distribution by individual control of the Peltier elements in the upper and lower row |
---|---|
Peltier | energy-saving Peltier heating-/cooling system integrated in the rear (heat pump principle) |
Standard equipment
Internals | 2 stainless steel grid(s), electropolished |
---|---|
Scope of delivery | incl. works calibration certificate for 10°C and 37°C |
Door | inner glass doors |
Door | fully insulated stainless steel doors with 2-point locking (compression door lock) |
Stainless steel interior
Volume | 749 l / 26.5 cu. ft. |
---|---|
Dimensions W x H x D in mm | w(A) x h(B) x d(C): 40.9 x 47.2 x 23.6 (d less 1.3” for fan – Peltier) |
Max. number of internals | 14 |
Max. loading of chamber | 440 lbs |
Max. loading per internal | 66 lbs |
Textured stainless steel casing
Installation | on lockable castors |
---|---|
Dimensions | w(D) x h(E) x d(F): 48.2 x 67.7 x 29.7 inches (d +2.2” door handle) |
Housing | rear zinc-plated steel |
Electrical data
Voltage Electrical load | 230 V, 50/60 Hz approx. 1300 W |
---|---|
Voltage Electrical load | 115 V, 50/60 Hz approx. 1300 W |
Ambient conditions
Set Up | The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm. |
---|---|
Ambient temperature | 16 ºC to 40 ºC |
Humidity rh | max. 70 %, non-condensing |
Altitude of installation | max. 2,000 m above sea level |
Overvoltage category | II |
Pollution degree | 2 |
Packing/shipping data
Transport information | The appliances must be transported upright |
---|---|
Customs tariff number | 8419 8998 |
Country of origin | Federal Republic of Germany |
WEEE-Reg.-No. | DE 66812464 |
Dimensions approx incl. carton | B x H x T: 1330 x 1910 x 1050 mm |
Net weight | approx. 230 kg |
Gross weight carton | approx. 301 kg |