Peltier-cooled Incubator IPP1060ecoplus
Temperature
Set temperature range in °C | from 0 (at least 20 below ambient temperature) to +70°C |
---|---|
Temperature | 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error |
Control technology
ControlCOCKPIT | TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-color displays. |
---|---|
Language setting | German, English, Spanish, French, Polish, Czech, Hungarian |
Timer | Digital backwards counter with target time setting, adjustable from 1 minute to 99 days |
Function HeatBALANCE | adapting the distribution of the heating performance of the upper and lower heating circuit from -50 % to +50 % |
Function SetpointWAIT | the process time does not start until the set temperature is reached |
Calibration | three freely selectable temperature values |
adjustable parameters | temperature (Celsius or Fahrenheit), fan speed, air flap position, program time, time zones, summertime/wintertime |
Ventilation
Fan | forced air circulation by 2 quiet air turbines, adjustable in 10 % steps for each segment individually |
---|---|
Fresh air admixture | adjustment of pre-heated fresh air admixture by air flap control in 10 % steps for each segment individually |
Vent | vent connection with restrictor flap |
Communication
Documentation | program stored in case of power failure |
---|---|
Programming | AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port |
Safety
Temperature control | mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 20°C above nominal temperature |
---|---|
Temperature control | overtemperature monitor TWW, protection class 3.1 or adjustable temperature limiter TWB, protection class 2, selectable on display |
AutoSAFETY | additionally integrated over- and under temperature monitor “ASF”, automatically following the setpoint value at a preset tolerance range, alarm in case of over- or under temperature, heating is switched off in case of overtemperature |
Autodiagnostic system | for fault analysis |
Alarm | visual and acoustic |
Standard equipment
Internals | 2 stainless steel grid(s), electropolished |
---|---|
Works calibration certificate | for +10°C and +37°C |
Door | fully insulated stainless steel doors with 2-point locking (compression door lock) |